Versatile, easy-to-use Trilogy 100 is designed for home, hospital and acute- care use to provide invasive and non-invasive ventilator support for adult and pediatric patients. A pressure and volume mode ventilator with proprietary algorithms, Trilogy 100 meets the changing needs of patients by combining unique flexibility with ease of use and advanced technology.

ข้อมูลผู้ติดต่อ

* ฟิลด์นี้เป็นฟิลด์ที่จำเป็น

ข้อมูลสำหรับติดต่อ

*
*
*

รายละเอียด บริษัท

*
*
*
*
*
*
*

รายละเอียดทางธุรกิจ

โปรดระบุความสนใจของคุณในการติดต่อเรา เพื่อการบริการอย่างดี ที่สุด
*
*

Since its launch in 2009, Trilogy has demonstrated clinical success and offered freedom and mobility to over 200,000 patients worldwide.

Specifications Show all

Dimensions and weight
Dimensions
16.68 cm L x 28.45 cm W x 23.52 cm H (6.6" L x 11.2" W x 9.3" H)
Weight
5 kg (11 lb)
Controls
Circuit types
Active with PAP; Passive
Ventilation types
Pressure and Volume
Volume modes
AC; SIMV (w/PS); CV; AC-MPV
Pressure modes
CPAP; S; S/T; T; PC; PC-SIMV (w/PS); PC-MPV; AVAPS-AE
Pressure
4 to 50 cm H₂O
IPAP
4 to 50 cm H₂O
PEEP/EPAP for active circuits
0 to 25 cm H₂O
PEEP/EPAP for passive circuits
4 to 25 cm H₂O
CPAP for passive circuits
4 to 20 cm H₂O
Pressure support differential
0 to 40 cm H₂O
Alarms
Circuit disconnect
Off; 5 to 60 sec
Apnea rate
Off; 10 to 60 sec
High tidal volume
Off; 50 to 2000 ml
Low tidal volume
Off; 40 to 2000 ml
High minute ventilation
Off; 1.0 to 99 l/min
Low minute ventilations
Off; 0.2 to 99 l/min
High respiratory rate
Off; 4 to 80 BPM
Low respiratory rate
Off; 4 to 80 BPM
Measured patient parameters
Respiratory rate
0 to 80 BPM
Minute ventilation
0 to 99 l/min
Vte/Vti
0 to 2000 ml
Leak rate
0 to 200 l/min
Peak inspiratory flow
0 to 200 l/min
Peak inspiratory pressure
0 to 99 cm H₂O
Mean airway pressure
0 to 99 cm H₂O
% Patient triggered breaths
0 to 100 %
I:E ratio
9.9:1 to 1:9.9
Controls
Tidal volume
50 to 2000 ml
Breath rate
0 to 60 BPM for AC mode; 1 to 60 BPM for all other modes
Inspiratory time
0.3 to 5.0 sec
Rise time
1 to 6
Ramp start pressure
0 to 25 cm H₂O for active circuits; 4 to 25 cm H₂O for passive circuits; 4 to 19 cm H₂O in CPAP
Ramp length
Off; 5 to 45 min
Flex
Off; 1 to 3
Flow trigger sensitivity
1 to 9 l/min
Flow cycle
10 to 90 %
Apnea rate
4 to 60 BPM